TY - JOUR N2 - In this study, a procedure for optimal selection of measurement points using the D-optimality criterion to find the best calibration curves of measurement sensors is proposed. The coefficients of calibration curve are evaluated by applying the classical Least Squares Method (LSM). As an example, the problem of optimal selection for standard pressure setters when calibrating a differential pressure sensor is solved. The values obtained from the D-optimum measurement points for calibration of the differential pressure sensor are compared with those from actual experiments. Comparison of the calibration errors corresponding to the D-optimal, A-optimal and Equidistant calibration curves is done. L1 - http://www.journals.pan.pl/Content/106341/PDF/10.15.15mms-2016-0029%20paper%2009.pdf L2 - http://www.journals.pan.pl/Content/106341 PY - 2016 IS - No 3 EP - 424 DO - 10.1515/mms-2016-0029 KW - sensor KW - calibration curve KW - calibration design KW - D-optimality criterion KW - least squares method A1 - Hajiyev, Chingiz PB - Polish Academy of Sciences Committee on Metrology and Scientific Instrumentation VL - vol. 23 DA - 2016.09.30 T1 - Sensor Calibration Design Based on D-Optimality Criterion SP - 413 UR - http://www.journals.pan.pl/dlibra/publication/edition/106341 T2 - Metrology and Measurement Systems ER -